Safety device and error code – Yamato Scientific BH300 Thermo-Elite User Manual
Page 29

26
Safety device and error code
The self-diagnosis function built into by the controller and the independent safety device from the
controller are installed in this machine. The table below shows the cause and the processing method
when the safety device operates. When the safety device operates, the error code is displayed to a
main display. Deal with it according to the processing method shown here.
Safety device
Display
Cause and processing method
Notes
Leak Breaker
No display
Please report to our customer service
section and confirm the cause.
Motor thermal protector
No display
Overheating of motor : Please report to
our customer service section.
Automatic overheat
protector.
No display
Heat of sample : Please decrease the
sample.
Sensor abnormality
detection
TROUBLE sign blinking
blinking
Disconnection of temperature sensor :
Please report to our customer service
section.
Triac short-circuit
detection
TROUBLE sign blinking
blinking
Short-circuit of Triac : Please report to
our customer service section.
Heater short-circuit
detection
TROUBLE sign blinking
blinking
Disconnection of heater : Please report
to our customer service section.
independent temperature
overheating prevention
device
TROUBLE sign blinking
blinking
Incorrect setting of independent
temperature overheating prevention
device : Please set correctly.
Heat of sample : Please decrease the
sample.
Defect of independent temperature
overheating prevention device : Please
report to our customer service section.
BH500 and
the machine
equipped with
independent
temperature
overheating
prevention
device
Main relay defect
detection
TROUBLE sign blinking
blinking
Breakdown of main relay : Please
report to our customer service section.
Refrigerator abnormality
detection
TROUBLE sign blinking
blinking
Dropout of the plug of refrigerator :
Please connect with the outlet in the
back of the control part.
The refrigerator is abnormal : Please
report to our customer service section.
BH200/300
Water level abnormality
detection
TROUBLE sign blinking
blinking
Decrease in water level : Add liquid to
reservoir.
Rise of water level (thermal
expansion) :
Decrease liquid from reservoir.
Defect of water level detector : Please
report to our customer service section.
The
equipment
goes back to
the normal state
automatically
when liquid
level becomes
normal.