Module Status and Input Data

Chapter 6

6Ć2

The complete bit/word description for the block transfer read from the module
is defined in Table 6.B.

Table 6.B

Bit/Word Description for the 1771-IXHR Input Module

Word

Bit

Definition

Word 1

Bit 00

Power-up bit is set to indicate that the module is waiting for its first

write block transfer

Bit 01

Out of range bit is set if one or more channel inputs are above or

below the range for which you configured the module

Bit 02

Real time sample time-out bit is set when the module updates an

input buffer with new data before the processor has read the previous

data. Monitor this bit only if you select real time sampling.

Bit 03

Not used

Bit 04

Low cold junction temperature bit is set when the cold junction

temperature is less than 0.0

o

C or 32.0

o

F.

Bit 05

High cold junction temperature bit is set when the cold junction

temperature exceeds 60.0

o

C or 140.0

o

F.

Bit 06

Dynamic clamp bit. Prevents rapid changes in data due to data

corruption over the opto-isolation barrier as a result of ESD, radiation

bursts, etc.

0 = feature active

1 = feature inhibited

Bit 07

EEPROM calibration values could not be read.

Bits 08-15

Not used

Word 2

Bits 00-07

Underrange bit for each channel is set to indicate an input is out of

range: bit 00 for channel 1 thru bit 07 for channel 8.

Bits 08-15

Overrange bit for each channel is set to indicate an input is out of

range: bit 08 for channel 1 thru bit 15 for channel 8. Also set for open

channel detection.

Word 3

Bits 00-07

Low alarm bit for each channel is set to indicate the input is less than

the low limit value you entered in the corresponding low alarm word

(BTW word 4, 6, 8, 10, 12, 14, 16, or 18): bit 00 for channel 1 thru bit

07 for channel 8.

Bits 08-15

High alarm bit for each channel is set to indicate the input has

exceeded the high limit value you entered in the corresponding high

alarm word (BTW word 5, 7, 9, 11, 13, 15, 17, or 19): bit 08 for

channel 1 thru bit 15 for channel 8.

Words 4-11

Input for channel 1 through 8 respectively in 0.1

o

C or 0.1

o

F resolution

for temperature and 10

µ

V or 1

µ

V resolution for millivolts.

Word 12

Cold junction temperature in 0.1

o

C or 0.1

o

F.

Bit/Word Descriptions

Advertising