Specifications

Number of Inputs

8, all of the same type or 4 each of 2 different types

I/O Chassis Location

Any single I/O module slot

Type of Input (Selectable)

Type B, Pt-30% Rh/Pt-6% Rh (320 to 1800

o

C)

Type E, chromel/constantan

(-270 to 1000

o

C)

Type J, iron/constantan

(-210 to 1200

o

C)

Type K, chromel/alumel

(-270 to 1380

o

C)

Type R, Pt/Pt-13% Rh

(-50 to 1770

o

C)

Type S, Pt/Pt-10% Rh

(-50 to 1770

o

C)

Type T, copper/constantan

(-270 to 400

o

C)

Millivolt

(-100 to +100mV dc)

Thermocouple Linearization

IPTS-68 standard, NBS MN-125

Cold Junction Compensation

Range: 0 to 60

o

C

Accuracy: +0.5

o

C

Temperature Scale (Selectable)

o

C or

o

F

Input Resolution

3.2328

µ

V

Display Resolution

0.1

o

C, 0.1

o

F; or 1.0

µ

V, 10

µ

V

Input Isolation

1000V peak between inputs, between input and common, and

between input and backplane connections

Common Mode Rejection

120dB at 60Hz, up to 1000V peak

Common Mode Impedance

Greater than 10 megohms

Normal Mode Rejection

60dB at 60Hz over +100mV

Input Overvoltage Protection

120V rms, continuous

Open Input Detection

Open input produces an overrange in less than 10 seconds

Input Connections

18-terminal wiring arm (Cat. No. 1771-WI)

Data Format

2's complement binary

Calibration Methods

Auto - Auto-calibration for offset and gain

Manual - Zero offset and gain adjustment for each channel via

programming terminal

Verify every six months for maintaining absolute accuracy

Processor Compatibility

PLC-3 or PLC-5 family processor using the 1771 I/O structure and

block transfer. (Not recommended for use with PLC-2 family

processors.)

Environmental Conditions

Operating Temperature:

Rate of Change:

Storage Temperature:

Relative Humidity:

0 to 60

o

C (32 to 140

o

F)

Ambient changes greater than 0.5

o

C per minute may temporarily

degrade performance during periods of change

-40 to 85

o

C (-40 to 185

o

F)

5 to 95% (without condensation)

Backplane Power Consumption

750mA @ 5V; 3.75 Watts maximum

Field Wiring Arm

Cat. No. 1771-WI

Keying

Between 20 and 22

Between 24 and 26