Calibrating the crimp tool – Rockwell Automation 2090-Series Circular-DIN Connector Kits, Flange Kits, and Crimp Tools User Manual

Page 11

Advertising
background image

Rockwell Automation Publication 2090-IN042A-EN-P - May 2011

11

Calibrating the Crimp Tool

1. Pull the battery compartment open and remove the battery.
2. Turn the adjustment wheel all the way clockwise (towards the –) until the

adjustment wheel stops turning and the reference pin is at REF.

3. Replace the battery and close the battery compartment until it snaps

closed.

4. Close the handles of the crimp tool to the endstops and hold the handles in

the closed position.

5. Turn the adjustment wheel counterclockwise (towards the +) to open the

crimp indenters to the size of the 2 mm calibration gauge.

6. Place the 2 mm calibration gauge into the crimp indenters.

Turn
Adjustment
Wheel
Clockwise

Reference
Pin at REF

Insert Battery into
Backside of Battery
Compartment

Insert 2 mm
Calibration
Gauge Into
the Indenters