Installing the Enwatch 5

Connectors

The Enwatch unit has 11 connectors labeled J1 to J11 as shown in Table 1.1.

The location of the connectors is shown in Figure 1.2 on page 4. The pin
outputs are listed in the following tables.

Table 1.1 Enwatch connectors

Connector

No. of Pins

Function

J1

16

Analog input channels 1–8

J2

16

Analog input channels 9–16

J3

3

External trigger 1 / event 1

J4

3

External trigger 2 / event 2

J5

3

External trigger 3

J6

3

External trigger 4

J7

5

Test points (service use only)

J8

8

Ethernet interface

J9

2

Supply voltage

J10

2

Spare

J11

9

Serial port (RS-232)

Table 1.2 J1 and J2 pin outputs

J1: Analog Inputs 1-8

J2: Analog Inputs 9-16

Signal

Pin No

Signal

Pin No.

CH 1 input
CH 1 ground

1
2

CH 9 input
CH 9 ground

1
2

CH 2 input
CH 2 ground

3
4

CH 10 input
CH 10 ground

3
4

CH 3 input
CH 3 ground

5
6

CH 11 input
CH 11 ground

5
6

CH 4 input
CH 4 ground

7
8

CH 12 input
CH 12 ground

7
8

CH 5 input
CH 5 ground

9
10

CH 13 input
CH 13 ground

9
10

CH 6 input
CH 6 ground

11
12

CH 14 input
CH 14 ground

11
12

CH 7 input
CH 7 ground

13
14

CH 15 input
CH 15 ground

13
14

CH 8 input
CH 8 ground

15
16

CH 16 input
CH 16 ground

15
16