4.7 Results
■■■■■■■■■■■■■■■■■■■■■■
140
■■■■■■■■
743 Rancimat
Induction time
The induction time is used for extrapolation. If the induction time has
been determined manually, this value will be used and not the automati-
cally determined induction time.
Stability time
The stability time is used for extrapolation.
Formula 1 for induction time
The result calculated with formula 1 is used for extrapolation.
Formula 2 for stability time
The result calculated with formula 2 is used for extrapolation.
Formula 3 for stability time
The result calculated with formula 3 is used for extrapolation.
Target temperature
Temperature, for which the selected result is to be converted. The conver-
sion is realized using the regression straight line (from the extrapolation).
Range
freely selectable ... °C
"Live graph" tab
On the Live graph tab, you can carry out the settings for displaying the
live graph.