Metrohm 743 Rancimat User Manual
Page 160

4.8 GLP functions
■■■■■■■■■■■■■■■■■■■■■■
152
■■■■■■■■
743 Rancimat
■
Click on the menu item Tools
▶ GLP test ▶ Conductivity... and
confirm with [Next >].
The following dialog window appears:
■
Select the option Use standard solution.
■
Under Conductivity of standard solution, enter the conductiv-
ity in µS/cm and confirm with [Next >].
c(KCl) = 1 mmol/L can be used as a standard solution. This is
derived from the optionally available conductivity standard
6.2301.060 (KCl 0.1 mol/L) through dilution with distilled water. The
conductivity of this diluted solution is:
Temperature Conductivity
18 °C
19 °C
20 °C
21 °C
22 °C
23 °C
24 °C
25 °C
127 µS/cm
130 µS/cm
133 µS/cm
136 µS/cm
138 µS/cm
141 µS/cm
144 µS/cm
147 µS/cm
The conductivity is measured sequentially at all channels.
After the last measurement is completed, the dialog window with
the test results appears.