4 carrying out glp tests, 1 temperature, Carrying out glp tests – Metrohm 743 Rancimat User Manual
Page 153: Figure 7, Accessories for glp test temperature

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
743 Rancimat
■■■■■■■■
145
4.8.4
Carrying out GLP tests
4.8.4.1
Temperature
For the GLP test for temperature measurement the calibrated tem-
perature sensor and the measuring insert are required (part of the option-
ally available GLP test set, see Optional accessories, page 185). For this,
the temperature sensor must be connected to the Pt100 connector on the
rear of the 743 Rancimat.
2
1
5
4
3
Figure 7
Accessories for GLP test Temperature
1
Pt100 temperature sensor
2
PTFE cylinder
3
Hexagon screw
4
Hexagon key (6.2621.120)
5
Aluminum cylinder
Preparing the GLP test set
Proceed as follows in order to prepare the GLP test set according to Figure
7:
1 Assemble accessories
■
Insert the temperature sensor from above into the opening of the
PTFE cylinder.
■
Place the PTFE cylinder on the aluminum cylinder in such a way
that the temperature sensor fits into this opening.
■
Push the temperature sensor into the aluminum cylinder as far as
the limit stop.