4 Operation

743 Rancimat

■■■■■■■■

147

2 Enter calibration points of the temperature sensor

■

Click on the menu item Tools

▶ GLP test ▶ Temperature...

and confirm with [Next >].
The following dialog window appears:

■

Enter a Sensor name for the temperature sensor.

■

Under Point 1…3, enter the Temperature and the Resistance
of the calibration point.

■

Confirm the entry with [Next >].

The following dialog window appears:

3 Enter test parameters

■

Under Channels used for test, select the required channels.
One channel per block has to be selected at least.

■

Under Test duration per channel, enter the required test dura-
tion and confirm with [Next >].