7 technical specifications, 1 general data, 2 temperature regulation and measurement – Metrohm 743 Rancimat User Manual
Page 178: General data, Temperature regulation and measurement

7.1 General data
■■■■■■■■■■■■■■■■■■■■■■
170
■■■■■■■■
743 Rancimat
7 Technical specifications
7.1
General data
Number of instru-
ments that can be
connected
1…4
Number of sam-
ples per instru-
ment
1…8 (4 each per heating block)
Sample amount
Several grams / several milliliters
7.2
Temperature regulation and measurement
Range setpoint
temperature
50…220 °C
Temperature cor-
rection (Delta T)
0…±9.9 °C
(can be entered manually or determined automatically using the exter-
nal temperature sensor)
Temperature
measuring range
0…250 °C
Resolution
0.1 °C
Max. deviation of
the heating block
temperature from
the set value
(50…220 °C)
±0.3 °C
Reproducibility of
the set tempera-
ture
typ. < 0.2 °C
Max. temperature
difference
between different
measuring posi-
tions per block
typ. < 0.3 °C
Temperature fluc-
tuations
typ. < 0.1 °C
(at attained operating temperature, inserted and identically filled reac-
tion vessels and 20 L/h air throughput)