4 Operation

743 Rancimat

■■■■■■■■

63

per block
The 4 channels of a block can be started and stopped jointly.

Stop criteria

Time

on | off (Default value: off)

Time until the automatic stop of the determination.

Range

0.1 ... 9999 h

Conductivity

on | off (Default value: off)

Conductivity limit value. If this value is reached, the determination is
ended automatically.

Range

1 ... 400 µS/cm

Endpoint(s)

on | off (Default value: off)

Automatic stop of the determination, as soon as every active channel has
reached all endpoints (induction time and stability time).

Table 2

Overview of stop criteria

Time

Conductivity

Endpoint(s)

Analysis stops...

not automatically.

✔

after the time entered, independently of reach-
ing the induction/stability time (endpoint).

✔

after reaching the conductivity entered, inde-
pendently of reaching the induction/stability
time (endpoint).

✔

after reaching the induction and stability time
(endpoint) defined on the Evaluation tab. If
the induction and stability time are evaluated,
both endpoints have to be found.

✔

✔

after reaching the first criterion (time or con-
ductivity), independently of reaching the induc-
tion/stability time (endpoint).

✔

✔

after reaching the first criterion (time or end-
point). If the time entered has been reached,
there is no further measuring until the induc-
tion/stability time (endpoint) has been reached.

Advertising