Metrohm 743 Rancimat User Manual

Page 61

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

743 Rancimat

■■■■■■■■

53

■

Click on the symbol .
The dialog window with the method parameters is opened.

■

Under Temperature enter the required setpoint temperature for
the sample and set Delta T to 0 °C.

■

Set the gas flow to the required value and finalize with [Save].

5 Start gas flow and heating

■

Click on the menu item Tools

▶ Gas flow control ▶ Gas flow

on.

■

Click on the button [Heat to xx °C].

Executing a Delta T determination

After you have prepared everything for determining Delta T (see "Prepar-
ing Delta T determination", page 52)
, you can now execute the determi-
nation of Delta T as follows:

1 Enter sensor calibration data

■

Click on the menu item Tools

▶ Determine Delta T ▶ Block A

or Block B.
The following dialog window appears:

■

Enter a Sensor name.

■

Enter the calibration data (temperature and resistance) for Points
1
to 3 and confirm with [OK].

Advertising