4 Operation

743 Rancimat

■■■■■■■■

149

5 Carry out measurement for further channels

Warning

The aluminum cylinder of the GLP test set is very hot. Do not
touch!

■

As soon as the message appears with the prompt to insert the
external temperature sensor together with the measuring insert in
the next selected channel, rapidly carry out the change.

■

Carry out measurement for each of the selected channels.

After the last measurement is completed, the dialog window with
the test results appears.

6 Confirm or reject results

✓

Test passed: The deviation of the measured temperatures
is within the specified tolerance.

✗

Test failed: The deviation of the measured temperatures is
greater than the specified tolerance.

■

Confirm the results with [Finish].