Metrohm 743 Rancimat User Manual
Page 157

■■■■■■■■■■■■■■■■■■■■■■
4 Operation
743 Rancimat
■■■■■■■■
149
5 Carry out measurement for further channels
Warning
The aluminum cylinder of the GLP test set is very hot. Do not
touch!
■
As soon as the message appears with the prompt to insert the
external temperature sensor together with the measuring insert in
the next selected channel, rapidly carry out the change.
■
Carry out measurement for each of the selected channels.
After the last measurement is completed, the dialog window with
the test results appears.
6 Confirm or reject results
✓
Test passed: The deviation of the measured temperatures
is within the specified tolerance.
✗
Test failed: The deviation of the measured temperatures is
greater than the specified tolerance.
■
Confirm the results with [Finish].
or
■
Reject the results with [Cancel].