4.8 GLP functions

■■■■■■■■■■■■■■■■■■■■■■

154

■■■■■■■■

743 Rancimat

■

Click on the menu item Tools

▶ GLP test ▶ Gas flow... and

confirm with [Next >].
The following dialog window appears:

■

Enter the Gas flow and confirm with [Next >].

The following dialog window appears:

2 Test all channels

■

Measure the gas flow sequentially for all channels and enter the
values in the corresponding fields K1…K4 in L/h.

■

After finishing the measurement, confirm the entries with
[Next >].

The dialog window with the test results appears.