6 Troubleshooting

743 Rancimat

■■■■■■■■

167

Problem

Cause

Remedy

The induction time
is longer/shorter
than expected.

The temperature is not
selected correctly.

■

Ensure that the correct method for the
determination has been selected.

■

Check that Temperature and Delta T
have been specified correctly in the
method (see Chapter 4.5.2.1, page 60).

See also: The induction
times are not reproducible
for multiple determina-
tions.

The stability time is
longer/shorter than
expected.

Delta Kappa is not defined
correctly.

Ensure that the value of Delta Kappa defined
in the method is correct (see Chapter 4.5.2.2,
page 64)
.

See also: The stability times
are not reproducible for
multiple determinations.

See also: The induction
time is longer/shorter than
expected.

The measurement
curves are extremely
noisy.

The air supply for the
measuring solution is direc-
ted to the conductivity
measuring cell.

Loosen tubing adapter (5-

11

) on the measur-

ing vessel cover, turn the PTFE tube (5-

17

) in

such a way that the air is no longer directed to
the measuring cell and fix in this position.

Gas bubbles adhere to the
conductivity measuring cell
during measurement.

■

Ensure that the measuring cell is clean and
free of fat. If necessary, clean thoroughly
(see Chapter 4.6.5, page 83).

■

In some cases ultra pure water contains a
large proportion of dissolved air which
escapes during measurement. In this case,
degas the ultra pure water before the
measurement for 5 to 10 min in a vacuum.

During the measurement,
sample evaporates in the
reaction vessel and con-
denses in the measuring
vessel. This results in con-
tamination of the conduc-
tivity measuring cell, which
in turn promotes adher-
ence of gas bubbles.

■

Keep the measuring time as short as possi-
ble, by about 4 to 6 h. The induction time
can be reduced by about half by increasing
the temperature by 10 °C.