6 "documentation" tab – Metrohm 743 Rancimat User Manual

Page 79

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

4 Operation

743 Rancimat

■■■■■■■■

71

AOCS Cd12b-92
Formula calculation according to standard AOCS Cd12b-92 with Tar-
get temperature
= 110 °C.
AOM
Formula calculation according to standard AOM with Target temper-
ature
= 97.8 °C.
free
Formula calculation with freely selectable Target temperature.

Formula parameters

Factor

Parameter for the calculation of the standard time, which indicates the
factor by which the time changes per 10 °C. This value can be determined
through the extrapolation (see Chapter 4.7.4, page 117) of results, which
have been recorded at different temperatures.

Range

0.001 ... 100

Target temperature

Target temperature for the calculation of the standard time.

Range

20 ... 300 °C

4.5.2.6

"Documentation" tab

This tab contains parameters for the automatic report output.