Data transfer – Metrohm VESUV User Manual

Page 105

Advertising
background image

100

• 7. Troubleshooting

Vesuv 3.0

Data transfer

Analytical devices

Problem

Remedy

Data are in the wrong device
window

If the devices are connected to the correct ports,
this is a hardware problem of your PC. Possible
remedies:

•

Switch on all devices (even if you don't need
them temporarily).

•

Use a Moxa or USB board to connect your de-

vices.

Garbled characters
On start of the datalogger

Check the handshake of the analytical device (you
may have software handshake there, whereas the
datalogger works with hardware handshake).

Garbled characters

•

Check the RS settings on your device and the
port parameter in the datalogger.

•

The analytical device has to send "IBM" charac-
ters (selection at the device in key <config>,
>peripheral units)

•

Don't send curves or reports which contain
curves from the analytical device.

•

Did you select the correct Device type in the
datalogger?

KF Coulometer: Missing
characters in raw reports

The KF Coulometer can't be connected to a Soft-
switch. Connect it directly to a PC COM or to a
Moxa or USB board resp.

Lock commands are not sent
to the device

•

The setting of the datalogger "Vesuv can send
data to device" has to be active ("Properties of
device window, Communication").

•

Remove the 6.2125.010 adapter from the RS
interface of your analytical device.

Method handling
(save/restore) is not possible

•

You can only save/restore methods for Titrinos,
756 KF Coulometer and 730 Sample Changer.

•

The setting of the datalogger "Vesuv can send
data to device" has to be active ("Properties of
device window, Communication").

•

Remove the 6.2125.010 adapter from the RS
interface of your analytical device.