2. Connecting devices
• 11
756 KF Coulometer
Cables
PC 25pin
6.2125.110
PC 9pin
6.2134.040
If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit, no parity, handshake "HWs".
The Coulometer cannot be connected via a Softswitch.
The RS parameters can be set on the Coulometer in key
<CONFIG>, >RS settings COM1.
Notes
•
Connect the PC to COM port 1 of the Coulometer.
•
The 756 KF Coulometer can't be connected to a Soft-
switch. Connect it directly to a PC port or to a port of an
extension board.
•
At the end of the titrations transmit at least the following
reports:
Result;Calc;MpList
Settings on the Coulometer under "<DEF>, >Report,
COM1"
You can also send additional reports which can be stored
in the database and/or printed out.
Important: Don't send any curves from the Coulometer
to Vesuv! The Vesuv curve is created from the measuring
point list which the Coulometer transmits.
•
If the Coulometer is operated by Vesuv then the report
identification, the date and method lines will be automati-
cally switched on in the report. This ensures that the re-
port is correctly recognized and read out.