712 conductometer – Metrohm VESUV User Manual
Page 20

Vesuv 3.0
2. Connecting devices
• 15
712 Conductometer
Cables
PC 25pin
6.2125.060
PC 9pin
6.2125.110
If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit and no parity. The handshake
must be set as follows:
•
HWs (if the Conductometer is connected to a COM port)
•
SWchar (if the Conductometer is connected via a Soft-
switch)
The RS parameters can be set on the Conductometer in key
<configuration>, >RS settings.
Settings on the Conductometer
•
For proper functioning you need the Conductometer pro-
gram version 712.0015 or higher.
•
Always transmit the print header:
settings under "<config>, >printer"
•
Switch off date/time transmission for each measurement:
settings under "<config>, >print meas.value"
•
Do not switch off sample number:
settings under "<config>, >auxiliaries"
•
Only transmit reports without embedded graphics, i.e.
short reports for cal T