• 7. Troubleshooting

Vesuv 3.0

Problem

Remedy

726 Titroprocessor:
Device window remains in
status "Data reception"

•

The interface of the Titroprocessor, which is
connected to the PC, can't be printer interface at
the same time (Setting at the Titroprocessor in
Config, Interfaces).

•

A curve was sent by the Titroprocessor. The end
of the report could not be recognized.

Stop the Titroprocessor

and restart it

.

•

After switching on the Titroprocessor it may have
sent garbled characters.

Stop the Titroprocessor

and restart it

.

See also device settings of 726 Titroprocessor,
page 9.

Balances

Problem

Remedy

No balance data in the bal-
ance window

•

Check the RS settings on your balance and the
port parameter in the datalogger.

•

The balance can't be connected to a Softswitch.
Connect it directly to a PC COM or to a Moxa or
USB board resp

•

Switch balance off/on.

•

Restart your PC.

•

Do you use the correct cable and is the cable
correctly plugged in ?

No balance data in the de-
vice. They are correct in the
balance window

•

Did you assign the balance to this device (see
status bar of device window)?

•

In the settings of the device, the menu item
"Device window, Properties, Communication,
Vesuv can send data to device" has to be active

in the datalogger.

•

With automatic assignment of balance: Did you
set the device for automatic balance assignment
also?

•

With automatic assignment of balance: Is your
analytical device in the sample size request?

•

Check if the correct balance type is set in your
analytical device (key <config>, >peripheral
units, for 726 Titroprocessor "Config, Inter-
faces").

•

Did you select the correct balance type in the
datalogger?

•

Remove the 6.2125.010 adapter from the RS
interface of your analytical device.