Vesuv 3.0

2. Connecting devices

• 9

726 Titroprocessor

Cables

PC 25pin

6.2125.060

PC 9pin

6.2125.110

If possible work with the following RS parameters: 9600
Baud, 8 Data Bit, 1 Stop Bit and no parity. The handshake
must be set as follows:

•

HWs (if the Titroprocessor is connected to a COM port)

•

SWchar (if the Titroprocessor is connected via a Soft-
switch)

The RS parameters can be set on the Titroprocessor in
"Config, Interfaces".

Notes

•

For working with Vesuv you need program version
5.726.0012 or higher.

•

Transmit at least the raw data report:
settings on 726 Titroprocessor under "Methods, *Report"
If you want to store reports, you should also transmit the
report "Methods, *Standard".
Important: Don't send any curves from the Titroproces-
sor to Vesuv! The Vesuv curve is created from the
measuring point list which the Titroprocessor transmits.

•

The report must be transmitted and not printed out.

•

Set data transfer to "LIMS":
settings under "Config, Interfaces"

•

If you are working with a balance which is connected to
the PC then you must first select the balance type on the
726 Titroprocessor.
Setting under "Config, Interfaces "

•

Only the first line of the report header can be read into
the database.

•

If sample data (sample identifications, weight) are altered
during the determination then they may differ in the indi-
vidual sub-determinations.

•

Equations with special expressions (TST) cannot be re-
calculated in Vesuv.