LaMotte SMART Spectro Spectrophotometer User Manual

Page 121

Advertising
background image

APPLICATION:

Domestic and industrial wastes.

RANGE:

0–150 mg/L COD

MDL:

5 mg/L

METHOD:

Dichromate in the presence of silver salts, at high

temperature in a closed system, oxidizes most organic

compounds to 95-100% of the theoretical amount. This

process is called digestion. As dichromate oxidizes

the organic compounds, the amount of yellow color

is reduced. The remaining yellow color is measured

colorimetrically at the 420 nm and is directly proportional

to the COD of the sample.

SAMPLE HANDLING

& PRESERVATION:

Collect samples in glass and test as soon as possible. If

samples must be stored, preservation is accomplished

by the addition of concentrated H2SO4 to adjust the

pH below 2. Samples with suspended solids should be

homogenized in a blender (100 mL for 30 seconds) and

then stirred gently with a magnetic stirrer.

INTERFERENCES:

Volatile organic compounds are not oxidized to the

extent that they are in the vapor above the digestion

solution. Therefore, they do not contribute to the

COD reading. Chloride concentrations above 10% of

COD interfere with the mercury free tubes. Chloride

above 2000 ppm will interfere with the mercury tubes.

Nitrite gives a positive interference of 1.1 ppm O

2

per ppm NO

2

–N which is insignifi cant unless nitrite

concentrations are very high. Other reduced inorganic

compounds are stoichiometrically oxidized, causing

a positive interference. Corrections can be made for

these compounds based upon their stoichiometry and

concentrations.

COD, Low Range

SMART Spectro Test Procedures 2.11

Test P

rocedures

Advertising