Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 181

Advertising
background image

PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

55 Manganese L) from TESTING MENU.

4. Scroll to and select 55 Manganese L from menu.
5. Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro. Use the 1.0 mL pipet (0354) to add 2.0 mL (two

measures) of *Hardness Buffer Reagent (4255). Swirl to mix.

8. Add 2 drops of *Sodium Cyanide, 10% (6565). Cap and mix.
9. Use the 0.5 mL pipet (0369) to add 0.5 mL of *Manganese Indicator Reagent

(3956). Cap and mix.

10. Immediately insert tube into chamber, close lid and select SCAN SAMPLE.

Record result.

11. Press OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

MANGANESE, Low Range

SMART Spectro Test Procedures 2.11

Test P

rocedures

Advertising