Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 75

Advertising
background image

PROCEDURE

Use universal sample holder

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Select

ALL TESTS (or another sequence containing 7 Biguanide from

TESTING MENU.

4. Scroll to and select 7 Biguanide from menu.
5. Rinse a tube (0290) with sample water. Fill to 10 mL with sample.
6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove the tube from colorimeter.
8. Use the 1.0 mL pipet (0354) to add 2.0 mL of Biguanide Indicator (3994). Cap

and invert three times to mix.

9. Wait 1 minute.
10. Insert the tube into chamber. Close lid.
11. Select SCAN SAMPLE. Record result in ppm Biguanide
12. Press OFF button to turn the spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

NOTE: For best possible results, a reagent blank should be determined to account

for any contribution to the test result by the reagent system. To determine the

reagent blank, follow the above test procedure to scan a distilled or deionized

water blank. Then follow the above procedure to perform the test on a distilled or

deionized water sample. This test result is the reagent blank. Subtract the reagent

blank from all subsequent test results of unknown samples. It is necessary to

determine the reagent blank only when a new lot number of reagents is obtained.

BIGUANIDE

SMART Spectro Test Procedures 2.11