LaMotte SMART Spectro Spectrophotometer User Manual

Page 28

Advertising
background image

28

SMART Spectro Operator’s Manual 2.11

5. Press

*ENTER to select ALL

TESTS.

ALL TESTS

*

1 Alkalinity-UDV
2 Aluminum
3 Ammonia-N L F
4 Ammonia-N L S

6. Press or

to scroll to the

desired test.

ALL TESTS

1 Alkalinity-UDV

*

2 Aluminum
3 Ammonia-N L F
4 Ammonia-N L S

7. Press

*ENTER to select the test.

The SMART Spectro is ready to

scan. The proper wavelength has

been selected.

2 Aluminum

*

SCAN BLANK
SCAN SAMPLE
END 535 NM

8. Insert the blank. Press

*ENTER

to scan the blank. The blank has

been stored.

2 Aluminum

SCAN BLANK

*

SCAN SAMPLE
END 535 NM

9. Insert the reacted sample. Press

*ENTER to scan the sample. The

result will be displayed.

2 Aluminum
T = 16.5%T
A = 0.7834A
C = 1.28 PPM
PRINT PRESS ENTER