SMART Spectro Operator’s Manual 2.11
5. Press
*ENTER to select ALL
TESTS.
ALL TESTS
*
1 Alkalinity-UDV
2 Aluminum
3 Ammonia-N L F
4 Ammonia-N L S
6. Press or
to scroll to the
desired test.
ALL TESTS
1 Alkalinity-UDV
*
2 Aluminum
3 Ammonia-N L F
4 Ammonia-N L S
7. Press
*ENTER to select the test.
The SMART Spectro is ready to
scan. The proper wavelength has
been selected.
2 Aluminum
*
SCAN BLANK
SCAN SAMPLE
END 535 NM
8. Insert the blank. Press
*ENTER
to scan the blank. The blank has
been stored.
2 Aluminum
SCAN BLANK
*
SCAN SAMPLE
END 535 NM
9. Insert the reacted sample. Press
*ENTER to scan the sample. The
result will be displayed.
2 Aluminum
T = 16.5%T
A = 0.7834A
C = 1.28 PPM
PRINT PRESS ENTER