LaMotte SMART Spectro Spectrophotometer User Manual

Page 6

Advertising
background image

6

SMART Spectro Operator’s Manual 2.11

SPECIFICATIONS

INSTRUMENT TYPE: Single beam spectrophotometer

Readout

5 line, 18 character per line LCD

Wavelengths

350-1000 nm

Wavelength Accuracy

± 2 nm

Wavelenth Resolution

1 nm

Wavelength Bandwidth

5 nm (max)

Photometric Range

0-125%T, –0.1-2.5A

Photometric Accuracy

± 0.005A

Photometric Stray Light <0.5%T
Dispersive Device

1200 Lines/mm ruled grating

Sample Chamber

Accepts 25 mm diameter fl at-bottomed test tubes,

10 mm square cuvettes, 16 mm COD test tubes

Source Lamp

Quartz halogen

Modes

%/T, ABS, pre-programmed tests

Pre-Programmed Tests

YES, with automatic wavelength selection

Languages

English or Chinese

User Defi ned Tests

Up to 25 user tests can be input

RS232 Port

8 pin mDIN

Power Requirements

Battery Operation (optional): Ni-Metal Hydride

battery pack Line Operation: 110/220V, 50/60 Hz

Dimensions

36 cm (wide) x 28 cm (deep) x 17 cm (tall)

Weight

10.3 lbs, 4.65 kgs

STATISTICAL AND TECHNICAL DEFINITIONS

RELATED TO PRODUCT SPECIFICATIONS

Method Detection Limit (MDL): “The method detection limit (MDL) is defi ned as

the minimum concentration of a substance that can be measured and reported

with 99% confi dence that the analyte concentration is greater than zero and is

determined from analysis of a sample in a given matrix containing the analyte.”

1

Note that, “As Dr. William Horwitz once stated, ‘In almost all cases when

dealing with a limit of detection or limit of determination, the primary purpose of

determining that limit is to stay away from it.’”

2

1. CFR 40, part 136, appendix B
2. Statistics in Analytical Chemistry: Part 7 – A Review, D. Coleman and

L Vanatta, American Laboratory, Sept 2003, P. 31.