PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

39 OXYGEN) from TESTING MENU.

4. Scroll to and select 39 OXYGEN from menu.
5. Rinse a clean tube (0290) with untreated sample water. Fill to the 10 mL line

with sample. This tube is the BLANK.

6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Fill a second tube (0290) to the 10 line with the treated “Fixed” sample. This

tube is the SAMPLE.

8. Remove BLANK from Spectro, insert SAMPLE tube into chamber, close lid and

select SCAN SAMPLE. Record result.

9. Press

OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.

SMART Spectro Test Procedures 2.11

DISSOLVED OXYGEN