PROCEDURE

Use universal sample holder.

1. Press and hold ON button until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

56 Manganese H) from TESTING MENU.

4. Scroll to and select 56 Manganese H from menu.
5. Rinse a tube (0290) with sample water. Fill to the 10 mL line with sample.
6. Insert tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from Spectro. Use the 0.1 g spoon (0699) to add two measures

of Manganese Buffer Reagent (6310). Cap and mix until powder dissolves.

8. Use the 0.15 g spoon (0727) to add one measure of *Manganese Periodate

Reagent (6311). Cap and shake for one minute. An undissolved portion of the

reagent may remain in the bottom of the tube without adversely affecting the

test results. Wait two minutes for maximum color development. Solution will

turn pink if manganese is present.

9. At the end of the two minute waiting period, mix, insert tube into chamber,

close lid and select SCAN SAMPLE. Record result.

10. Press OFF button to turn spectrophotometer off or press EXIT button to exit to

a previous menu or make another menu selection.