LaMotte SMART Spectro Spectrophotometer User Manual
Page 43
SMART Spectro Operator’s Manual 2.11
43
15. Press
EXIT to return to the
SELECT DEGREES menu. Press
EXIT again to return to INPUT
DATA NUMBERS. Press EXIT
again to return to the
USER TEST
menu.
102 H2O
*
NAME THE TEST
SELECT WL
NEW CALIBRATION
FORMAT RESULT
SELECTING THE NUMERICAL FORMAT OF THE
RESULT
To input tests with very different ranges, the number of decimal places displayed
for a result can be selected. A test which ranges from 20 to 1000 ppm should
not be displayed with three decimal places. A test with a range from 0.010 to
0.500 needs three decimal places (the microprocessor will always calculate the
concentration to many more signifi cant fi gures than will be displayed). Menu
choices of 0, 1, 2, or 3 decimal places will be given for the display.
1. Press to scroll to
FORMAT
RESULTS.
102 H2O
NAME THE TEST
SELECT WL
NEW CALIBRATION
*
FORMAT RESULT
2. Press
*ENTER to select FORMAT
RESULTS.
DECIMAL PLACES?
0 PLACES
1 PLACE
2 PLACES
3 PLACES
3. Press or
to select the
number of decimal places to be
displayed.
DECIMAL PLACES?
0 PLACES
1 PLACE
*
2 PLACES
3 PLACES