Procedure – LaMotte SMART Spectro Spectrophotometer User Manual

Page 133

Advertising
background image

PROCEDURE

Use universal sample holder.

1. Press and hold ON burton until spectrophotometer turns on.
2. Scroll to and select PROGRAMMED TESTS.
3. Scroll to and select ALL TESTS (or another sequence containing

28 Color) from TESTING MENU.

4. Scroll to and select 28 Color from menu.
5. Rinse a tube (0290) with color-free water (distilled or deionized water). Fill to 10

mL line with color-free water.

6. Insert the tube into chamber, close lid and select SCAN BLANK.
7. Remove tube from spectrophotometer. Empty tube.
8. Rinse tube with sample water. Fill to 10 mL line with water sample.
9. Insert tube with sample water, close lid and select SCAN SAMPLE. Record

result in color units.

10. Press OFF button to turn the spectrophotometer off or press EXIT button to

exit to a previous menu or make another menu selection.

COLOR

SMART Spectro Test Procedures 2.11