0 system installation – Bio-Rad EXQuest Spot Cutter User Manual

Page 14

Advertising
background image

2.0

SYSTEM INSTALLATION

Before the EXQuest

spot cutter can be used, the software must be installed and the instrument set up and

calibrated. At the time of installation, a Bio-Rad service representative will setup the instrument, install
PDQuest

and/or Quantity One

®

software and then calibrate the instrument for you.

2.1

PRE-INSTALLATION REQUIREMENTS

Before installing the EXQuest spot cutter, ensure that the following requirements have been met.

Installation Requirements Checklist

Bench Space

The EXQuest spot cutter should be placed on a sturdy lab bench. Tables are
not recommended but may be used if they are made of heavy materials and
are anchored to the floor. The EXQuest spot cutter requires a bench top area
that is 36 inches (91 cm) wide, 27 inches (69 cm) deep and 26 inches (66 cm)
high. There should be 38 inches (96 cm) of clearance above the front of the
bench to allow the spot cutter door to open upward.

There must be at least 10 inches of free space on the left side of the spot
cutter for access to the water and waste bottles.

There must be at least 12 inches of free space on both the left and right
side of the spot cutter during installation.

Space for a computer with monitor.

Power

A dedicated 20 Amp electrical circuit and two receptacles

Nominal input voltage range 100 V to 240 V (50 Hz/60 Hz)

Power consumption 220 W

Surge protector Or Uninterruptible Power Supply (UPS), recommended

Computer

Minimum Requirements:

Processor 1.2 GHz or greater

USB 1.1 or 2.0

2.1 GB hard drive

RAM 512 MB or greater

Screen Resolution 1,024 x 768 pixels

2 MB video RAM

Operating system Windows 2000 or Windows XP

Environment

Operating Temperature:

15–30°C·

Humidity

15–85%, non condensing

2.2

UNPACKING THE EXQUEST SPOT CUTTER

The EXQuest spot cutter and its accessories are shipped together in a single palleted crate. Before
unpacking the EXQuest spot cutter, carefully read the unpacking instructions located on the outside of the
packaging. Inspect the container for any damage that may have occurred in shipping. Severe damage to a
container may indicate damage to its contents. If you suspect damage may have occurred, immediately file a
claim with the carrier in accordance with their instructions before contacting Bio-Rad Laboratories.

SYSTEM INSTALLATION

2-1