Figure 3-10. Plate position calibration: plate selection.

b.

Use the Left/Right and Front/Back slider to center the cutting tip over well A1, as shown below
in Figure 3-11. Click on the slider bar to move the tip in 0.2 mm increments or click and drag the
slider to make larger adjustments.

c.

When the tip is centered, check Lower head to adjust eject height to lower the cutting tip into
the well. Use the sliders to center the tip in the well.

Figure 3-11. Plate position calibration: cutting tip adjustment.

EXQUEST CALIBRATION WIZARD

3-8