Bio-Rad EXQuest Spot Cutter User Manual

Page 32

Advertising
background image

Figure 3-13. Wash station position calibration.

b.

When the tip is centered, check Lower head to adjust wash height to lower the cutting tip into
the well. Use the sliders to center the tip in the well.

c.

Use the Raise/Lower slider to adjust the tip height. Typically the gel tip should be placed about
2 mm into its wash port and the Membrane cutting tip is about 2 mm or more into its wash port.
Note that the membrane cutting tip must be in contact with the water in the microtube in order to
effectively clean the tip.

d.

Click Next to save the wash port calibration, Back to return to the previous step or Cancel to
abort the calibration.

12.

Calibration Complete. The calibration complete step displays the Calibration Summary for the
calibration steps performed. Click Finish to exit the Calibration Wizard.

EXQUEST CALIBRATION WIZARD

3-10