3.8 Run the Pro260 Analysis

1. Open the lid of the electrophoresis station by pulling the release latch. Place the primed and loaded

chip on the chip platform and close the lid.

2. In the Experion software toolbar, click New Run

. In the New Run screen (Figure 3.4), from

the Assay pull-down list, select Protein > Protein 260.

3. Either select a project folder for the run from the Project pull-down list or create a new project folder

by entering a name in the Project field or by selecting File > Project > New. The project folder
appears in the project tree.

4. Enter a name for the run in the Run Prefix field and click Start Run

.

5. In the New Run dialog (Figure 3.5), select the number of samples to be analyzed. Though all wells

are filled, the Experion system stops the analysis when it reaches the number of samples entered.

Fig . 3 .5 . New Run dialog . The Experion system stops analysis when it
reaches the number of samples entered.

Fig . 3 .4 . Details of the New Run screen . The green dot in the lower right corner indicates that communication
between the electrophoresis station and Experion software has been established.