Bio-Rad Experion Protein Analysis Kits User Manual

Page 61

Advertising
background image

Table C.9. Detergents and reductants with severe effects on Experion Pro260 separation or analysis.

Detergent/Reductant

Concentration

Effect(s)

ASB-14

2%

Enlarged system peak shifts to right; 10 kD peak is missing

n-Dodecyl-

b-D-maltoside

1%

Enlarged system peak shifts to right; 10 kD peak is missing

NP40

2%

Enlarged system peak shifts to right; 10 kD peak is missing

SB3-10

2%

Enlarged system peak shifts to right; 10 kD sizing affected

Triton X-100

1%

Enlarged system peak shifts to right; 10 kD peak is missing

Tween-20

1%

Enlarged system peak shifts to right; 10 kD peak is missing

Tween-80

1%

Enlarged system peak shifts to right; 10 kD peak is missing

Zwittergent 3-14

1%

Enlarged system peak shifts to right; 10 kD peak is missing

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

57

Experion Pro260 Analysis Kit