Bio-Rad Experion Protein Analysis Kits User Manual

Page 60

Advertising
background image

Table C.5. Chemicals with mild effects on Experion Pro260 separation or analysis.

Chemical

Concentration

Effect(s)

Acetone

15%

Quantitation affected

Bio-Lyte

®

3/10

0.75%

Quantitation affected

Guanidine HCl

25 mM

Enlarged system peak shifts to right; 10 kD sizing affected

KCl

1.5 M

Quantitation affected

MES

50 mM, pH 6.0

Quantitation affected

MOPS

100 mM, pH 7.2

Quantitation affected

PEG

0.50%

Quantitation affected

Propylene glycol

50%

Enlarged system peak shifts to right; 10 kD sizing affected

Sucrose

200 mM

Quantitation affected

Table C.6. Chemicals with severe effects on Experion Pro260 separation or analysis.

Chemical

Concentration

Effect(s)

Ammonium sulfate

1.5 M

Protein precipitation

Guanidine isothiocyanate

1 M

Protein precipitation; no markers appear

Guanidine HCl

1 M

Protein precipitation; no markers appear

HCl

150 mM

Protein precipitation; markers missing

Pluronic F-68

0.1%

Protein smears

Detergents and Reductants

Table C.7. Detergents and reductants with no observable effects on Experion Pro260 separation or analysis.

Detergent/Reductant

Concentration

ASB-14

0.10%

CHAPS

0.50%

CHAPSO

0.10%

Deoxycholate

1.0%

Octyl-thioglucopyranoside

0.50%

Tributylphosphine (TBP)

7.5 mM

Triton X-100

0.40%

Tween-20

0.50%

Table C.8. Detergents and reductants with mild effects on Experion Pro260 separation or analysis.

Detergent/Reductant

Concentration

Effect(s)

ASB-14

0.5%

Enlarged system peak shifts to right; 10 kD sizing affected

CHAPS

1%

Enlarged system peak shifts to right; 10 kD sizing affected

CHAPSO

0.2%

Enlarged system peak shifts to right; 10 kD sizing affected

Deoxycholate

2%

Enlarged system peak shifts to right; 10 kD sizing affected

LDS

3%

Enlarged system peak shifts to right; 10 kD sizing affected

NP40

0.5%

Enlarged system peak shifts to right; 10 kD sizing affected

SB3-10

0.1%

Enlarged system peak shifts to right; 10 kD sizing and quantitation

affected

SDS

1%

Quantitation affected

Tween-20

0.75%

Enlarged system peak shifts to right; 10 kD sizing affected

Zwittergent 3-14

0.10%

Enlarged system peak shifts to right; 10 kD sizing affected

β-Mercaptoethanol

1.5 M

Quantitation affected

DTT

150 mM

Quantitation affected

56

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

Experion Automated Electrophoresis System