5 changing peak finding parameters, 6 changing general settings – Bio-Rad Experion Protein Analysis Kits User Manual

Page 41

Advertising
background image

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

37

6.5 Changing Peak Finding Parameters

The parameters entered tell the peak-find algorithm whether a peak is significant.

To edit the parameters used by Experion software for finding peaks in a selected sample/well:

1. Select the sample by clicking on the lane in the virtual gel (or on the sample name in the tree view).

2. Under the Settings tab, edit the parameters under Peak Find Settings:

n

Slope Threshold — represents the variation in fluorescence units over time required to

detect a peak (accepted values are 0.2–1,000, with lower values yielding more peaks)

n

Min . Peak Height — minimum height required for a peak to be integrated (accepted values

are 0.1–1,000). Determine the appropriate value for this parameter by zooming in on a small
peak in an electropherogram and reading its fluorescence value on the y-axis

n

Min . Peak Width — minimum amount of time that must elapse before a peak is

recognized (accepted values are 0.1–10, with lower values yielding more peaks)

3. Click Apply to apply the changes and reanalyze the data. Click Save in the main screen toolbar to

save the new conditions or click Reset to Default to recover the default settings.

To modify the peak-find settings for all sample wells:

1. Under the Settings tab, click the All Well Settings tab or select Analysis > All Wells Settings.

2. In the All Wells Settings dialog, expand Peak Find Settings and select the type of modification you

would like to apply (see above for descriptions).

3. Click OK to apply changes and Yes to overwrite the settings. Click Save in the main screen toolbar

to save the new conditions. To reset to default settings, click Reset to Default.

Do not adjust the Peak Find Settings values so that the markers are eliminated from analysis.

6.6 Changing General Settings

These settings appear in the Settings tab, in the All Wells Settings dialog under General Settings
(Figure 6.1). They apply to all samples in a run.

n

Data Frequency — displays the rate of data collection in Hz

n

Display Start Time — displays the time Experion software begins displaying peaks

n

Display End Time — displays the time Experion software stops displaying peaks

n

Upper Marker Concentration — displays the concentration of the upper marker

n

Use Time Corrected Areas — corrects for peaks of different sizes that pass the detector

at different rates

n

Calibrate All Valid Peaks — select this option if you are running a calibration curve and

want Experion software to calibrate all peaks (this is the same as selecting Quantitate All
Proteins in the Internal Std . Protein Calibration Curve window)

n

Filter Width — determines the width of the polynomial (in sec) to be applied to the data for

filtering (noise reduction). Set this variable to less than twice the width of peaks. Valid values
are 0.3–10

n

Polynomial Order — defines the power series applied to fit the raw data. Valid values are

1–10, with lower values yielding smoother curves

Experion Pro260 Analysis Kit

Advertising