Bio-Rad Experion Protein Analysis Kits User Manual

Page 49

Advertising
background image

Error

Probable Cause

Recommended Action

Protein quantitation is
incorrect

Upper and lower markers were not properly
assigned by the software

Check that both markers were properly assigned. Manually
select the marker(s), if necessary (follow the instructions in
Section 6.3, Manually Setting a Marker)

Pipetting and/or dilution errors occurred during
sample preparation or chip loading

Ensure your calculations and dilutions are correct

Ensure that pipets are calibrated, and use pipets that
accurately deliver volumes of 10 µl or less

Do not modify the sample preparation and loading protocols
described in this manual

Ensure that the total protein concentration of your samples
(including any user-defined internal standards that may be
added) is within the linear dynamic range before proceeding
with the sample preparation instructions described in
Section 3.4; if total protein concentration is unknown, run a
Pro260 analysis of a dilution series of your sample

Review the essential practices described in Chapter 2
before initiating another analysis with a new chip

Samples are old or have not been stored
properly, and the proteins degraded

Prepare fresh samples and try the analysis again

Samples were not denatured properly;
regardless of whether reducing or nonreducing
conditions are used, the Pro260 ladder and
samples must be heat-denatured for the assay
to run properly

Heat the samples and the Pro260 ladder for 3–5 min at
95–100°C before loading them into the chip. Keep the
caps of the sample and ladder tubes closed during heating
to prevent loss of vapor and subsequent changes in
concentration. Briefly centrifuge the samples before opening

Samples do not have the same staining
properties as the upper marker

Use a different internal standard or a calibration curve
(see Chapter 5, Protein Quantitation Methods, for more
information)

Ghost peaks or
contaminants appear in
electropherograms

Electrodes are contaminated

Follow the deep cleaning procedure described in Appendix
B and the software Help section (search term “electrodes”)

Use only 0.2 µm-filtered water (such as ReadyPrep
proteomics grade water) for diluting the protein samples and
Pro260 ladder; do not use autoclaved water

Upper or lower marker is
incorrectly identified

Sometimes Experion software selects the
incorrect peak to correspond to the upper or
lower marker. Such an error affects alignment
and sizing

Manually select the upper and lower markers by following
the instructions in Section 6.3

Portion of the
electropherogram is
incorrectly identified as a
peak

Default integration settings are inapporpriate
for the sample

Manually exclude a peak (see Section 6.4) or change the
peak finding parameters (see Section 6.5)

Virtual gel does not
resemble an SDS-PAGE
separation of the same
samples

The gel view is a direct interpretation of
the electropherogram data. By default, the
fluorescence intensity of the bands in each
lane is scaled to the intensity/peak height
of the highest peak in that sample (scale to
local). Consequently, each gel lane may have
a different scale of fluorescence intensity. In
addition, Experion Pro260 analysis differs from
SDS-PAGE in several ways that may impact
separations

Adjust scaling of the virtual gel to Scale To Global
(see Section 4.2). This also adjusts the scale on all the
electropherograms. Adjust the intensity bar to the right of
the virtual gel as needed

In an Experion Pro260 analysis, the separation of the 10 kD
and 260 kD markers is almost linear, whereas in SDS-PAGE
it is logarithmic. To mimic the appearance of an SDS-PAGE
separation, select Tools > Options > Advanced and select
the Gel Mobility Correction option

Glycoproteins are not
separating as expected

Large and complex glycosolated proteins may
not migrate as expected with the Experion
Pro260 assay; however, their separation
characteristics will be reproducible

For more information, refer to Bio-Rad bulletin 5453 or
contact Technical Support

Technical Support: 1-800-4BIORAD • 1-800-424-6723 • www.bio-rad.com

45

Experion Pro260 Analysis Kit

Advertising