4.2 TUBING INSTALLATION

1. Referring to Figure 4-2, pull the platen cam lever away from the

pump head to unlock the platen and slide the platen away from
the pump head frame assembly, exposing the rollers.

2. Slip a lock-ring onto one end of the tubing. See below for the size

and color of each lock-ring. Insert a barbed female luer fitting into
the same end until the tubing reaches the flange of the fitting.
Clamp the luer-fitting into place by sliding the lock ring along the
tubing over the barbed fitting (see Figure 4-3). Repeat on the
other end of tubing.

Note: Use of lock-rings is required only when operating at
pressures above 10 psi.

Tubing ID

Lock-Ring Color

0.51 mm (.02”)*

None Required

0.8 mm (1/32”)

Red

1.6 mm (1/16”)

Orange

3.2 mm (1/8”)

Yellow

*

The 0.51 mm ID tubing is placed inside the 3.2 mm ID tubing, thus no lock-

rings are required for the smaller tubing.

Figure 4-3. Attachment of tubing fittings.

Tubing Selection