Appendix b connection to other instruments – Bio-Rad Econo Gradient Pump Kits User Manual

Page 45

Advertising
background image

APPENDIX B
CONNECTION TO OTHER INSTRUMENTS

The I/O connector on the rear panel of the pump is for the operation
of the fraction collectors and chart recorder (see Figure 3-2). Output
signals are TTL compatible. To use these signals for non-Bio-Rad
devices, you must insure that the circuit external to the Econo
Gradient Pump does not draw more than 15 milliamperes of current.
The following table describes the rear panel mini-DIN connector
pinouts for this I/O connector.

Table B-1. I/O mini-DIN Connector Pinouts

Selected Fraction Collector

Pin#

Name

None

2110

2128

1

PS

held HIGH

held HIGH

held HIGH

starts 1327

starts 1327

starts 1327

2

RUN

---

2.5 s

held LOW

LOW PULSE starts 2128
resets 2110

3

FCNTRL1

held LOW

held LOW

held LOW

starts EGP

starts EGP

starts EGP

program

program

program

4

START

---

120 ms LOW

held LOW

PULSE

starts 2128

toggles between
start and stop

Other Instruments

41

PIN #7

PIN #8

PIN #5

PIN #2

PIN #4

PIN #6

PIN #3

PIN #1