Problem

Possible Cause

Solution

Flow rate not

Pump tubing

Replace tubing.

consistent

damaged or
worn out

Platen adjustment Adjust the platen by turning
incorrect

it counter clockwise until it is
completely loosened, then
rotate three to five full turns
clockwise. See section 4.3,
Platen Adjustment.

Wrong tubing size Refer to Figure 4-1 for

selection

selected

criteria.
Select proper ID tubing within
Setup.

Large change in

Re-calibrate flow within Setup.

back pressure or
fluid viscosity

Restriction on inlet Detach the inlet tubing and
side

Gradient Mixer. Rinse with
deionized water.

Excessive

Platen too tight

Adjust the platen by turning

wear on tubing

it counter clockwise until it is
completely loosened, then
rotate three to five full turns
clockwise. See section 4.3,
Platen Adjustment.

Troubleshooting