Bio-Rad Econo Gradient Pump Kits User Manual
Page 40

Problem
Possible Cause
Solution
Flow rate not
Pump tubing
Replace tubing.
consistent
damaged or
worn out
Platen adjustment Adjust the platen by turning
incorrect
it counter clockwise until it is
completely loosened, then
rotate three to five full turns
clockwise. See section 4.3,
Platen Adjustment.
Wrong tubing size Refer to Figure 4-1 for
selection
selected
criteria.
Select proper ID tubing within
Setup.
Large change in
Re-calibrate flow within Setup.
back pressure or
fluid viscosity
Restriction on inlet Detach the inlet tubing and
side
Gradient Mixer. Rinse with
deionized water.
Excessive
Platen too tight
Adjust the platen by turning
wear on tubing
it counter clockwise until it is
completely loosened, then
rotate three to five full turns
clockwise. See section 4.3,
Platen Adjustment.
Troubleshooting
36
4006161B.qxd 8/14/2003 8:09 AM Page 36