Bio-Rad Econo Gradient Pump Kits User Manual
Page 41

Problem
Possible Cause
Solution
Cannot
Pump must be
Insure that pump is stopped.
calibrate
stopped in order to
pump
calibrate pump
RPM has been
Select Flow Rate in screen 1 of
selected
Setup.
Valves do not Pump is off
Turn on the pump.
respond or
actuate
Valve cabled to
Mixer and/or Aux valve must be
wrong connector
plugged in to its correct socket.
Chart recorder Pump is off
Turn on the pump.
or fraction
collector does Valve cabled to
Chart recorder or fraction
not respond
wrong connector
collector must be plugged in to
its correct connector.
Wrong fraction
Setup screen 14 allows choice
collector specified of a Model 2110, 2128 or none.
in setup
Select the appropriate choice.
The proper
Valves plugged
See section 3.2 to plug valves
valves do not into wrong port.
into the correct port.
appear on the
display
EGP program Wrong device
Ensure that only the fraction
starts
plugged into I/O
collector or chart recorder is
unexpectedly port
plugged into the I/O port.
Ensure that pin #3 is not
shorted.
Troubleshooting
37
4006161B.qxd 8/14/2003 8:09 AM Page 37