The Econo Gradient Pump can be connected to the BioLogic HR or
BioLogic Duo-Flow workstation via an Instrument bus cable (System
cables 17, 18, 19, or 21). When connected, the address of the Econo
Gradient Pump must correspond to the address recognized by the
BioLogic System. This address is only relevant when the Econo
Gradient Pump is connected to the BioLogic System. The default
address is 1.

9 Address #

001

Use the arrow keys to enter a new address, and press Next to accept
and advance to screen 10.

Screen 10. Revolutions Counter.
The performance of the peristaltic tubing degrades with use. This
revolution counter provides a measure of wear experienced by the
tubing. The relation between revolutions and tube life must be
empirically determined because it is dependent on several factors
including load, tubing type and platen pressure.

10 Revolutions=

Reset ___

Pressing the Reset softkey resets the counter back to 0. Pressing
Next will accept the displayed value and advance to screen 11.

Screen 11. LCD Contrast.
The LCD display contrast is set here.

11 LCD Contrast

__%

Use the arrow keys to enter the desired contrast and press Next to
advance to screen 12.