Model 2110 Fraction Collector (continued)

When a programmed run is started, the EGP will
automatically start the fraction collector. If the SV-3
diverter valve is connected to the EGP, then fraction
collection will begin at the point in the program when
the diverter valve is in the FRAC position.

If HOLD is selected during the program, then the
pump will continue pumping at the current conditions
and fraction collection will continue as programmed.

If PAUSE is selected during the program, then the
pump and fraction collector will stop.

The pump and fraction collector will resume when
CONTINUE is selected.

Model 1327 Chart Recorder

As described above, these cables connect fraction
collectors and the Model 1327 Chart Recorder to the
EGP. The fraction collector part of the cable should
be put aside if a fraction collector is not present. The
EGP will automatically start and stop the chart
recorder. The parameters should be set at the chart
recorder itself.

If a fraction collector is present, then the fraction
collector marks are superimposed on the UV trace
directly.

If HOLD is selected during the program, then the
pump will continue pumping at the current conditions
and the chart recorder will continue recording.

Other Instruments

28

Either
Cable #22
(Catalog #
731-9010)
or
Cable #23
(Catalog #
731-9009)

4006161B.qxd 8/14/2003 8:09 AM Page 28