Bio-Rad Foresight™ Chromatography Columns, Prepacked User Manual

Page 23

Advertising
background image

18 Technical Support: 1-800-424-6723 • www.bio-rad.com

For instance, compared to a buffer solution,

a protein solution of relatively high protein

concentration will be prone to not forming clean

single droplets on the tip of the drip director.

The droplets tend to stick to the outer surface of

the drip director, forming even bigger droplets,

which may end up in multiple wells. In contrast,

centrifugation does not have these shortcomings

and is the preferred method for sample

collection. Using a media plate with

0.8 ml well volume loaded with 0.3 ml of

liquid and equipped with a 0.45 μm filtration

membrane, a centrifugation cycle at 300 x g

for 60–120 seconds should lead to complete

sample collection. However this should be

confirmed visually, since times up to 10 minutes

are occasionally required. A swinging bucket

rotor centrifuge, equipped with a microtiter plate

bucket is recommended.

For steps such as plate equilibration, vacuum

filtration can be used. Vacuum filtration for

1 minute (or until the wells are empty) at

–1 to –3” in Hg is recommended. Visual

inspection of the wells should be conducted prior

to sample loading.

3.3.5 Mixing

An orbital shaker capable of mixing at 1100 rpm

is recommended for good mixing. The shaker

should be equipped with plate holders. The

collection and media plate should be secured

to each other and to the microtiter plate shaker

before the start of the mixing cycle. The top of

the media plate should be sealed.

Advertising