Bio-Rad Foresight™ Chromatography Columns, Prepacked User Manual

Page 27

Advertising
background image

22 Technical Support: 1-800-424-6723 • www.bio-rad.com

3.5 Equipment and Materials Required for HT
Experiments

In addition to the Foresight chromatography media

plates the following equipment is required:

(a) 96-well collection plates, (b) 8- or 12-channel

pipets, (c) microplate shaker that can mix at

1100 rpm, (d) bucket centrifuge equipped with a

rotor for microtiter plates, and (e) microtiter plate

spectrophotometer.

HT experimentation analytics must accommodate

and be optimized for a large number of samples of

relatively small volume. Accessibility to an HPLC

system equipped with a microtiter plate autosampler

will increase the analytical throughput.

3.6 Detailed Experimental Protocol

3.6.1 Media Plate Preparation

a) Resuspend the medium to guarantee that

media adhered to the plate’s top seal returns

to the well. Either one of the following two

methods may be followed to ensure proper
resuspension of the media:
• Gently tap filter plate on benchtop 20

times; or

• Centrifiguge plate at 100 x g for 60–120

seconds

b) Remove the media plate’s bottom seal and

place the media plate on top of its

corresponding 96-well collection plate. Next,

peel off the top seal while holding the media

plate firmly.

c) Because the liquid storage solution does

not contain valuable solutes, remove the

solution by either vacuum filtration or

centrifugation.