Bio-Rad Foresight™ Chromatography Columns, Prepacked User Manual
Page 26
Advertising
Technical Support: 1-800-424-6723 • www.bio-rad.com 21
0
20
40
60
80
100
120
140
160
180
0.00 2.00 4.00 6.00 8.00 10.00
q
(m
g/
m
L_m
edi
a)
c (mg/mL)
Data
Langmuir model
(A)
0
20
40
60
80
100
120
140
160
180
200
0.00
1.00
2.00
3.00
4.00
q (
m
g/
m
L_
m
edi
a)
c (mg/mL)
Data
Langmuir model
(B)
Fig. 5. Application of error analysis to HT
experimentation isotherm data.
A, isotherm data on bovine serum albumin
(BSA)-Nuvia Q-pH 8.5, q
max
= 160 ± 12.3 mg/ml media;
K = 1.80 ± 0.80 mg/ml; B, human immunoglobulin G
(hIgG)-Nuvia-S-pH 5.0. q
max
= 208 ± 11.7 mg/ml media;
K = 0.82 ± 0.09 mg/ml.