Salter Brecknell 650 User Manual

Page 25

Advertising
background image

25

ITEM

SOFT

KEY

DESCRIPTION

Calibration of 650 or remote platform

CALIB

Allows you to enable or disable the

controlled accuracy feature

Zero Count Value/ re-entry

KCalZR

Allows re-entry of zero point value without

going through calibration procedure

Full scale count value/ re-entry

KcalSP

Sets the amount of filtering on the TC2010

platform

Allows setting and changing of 650

parameters

LOCAL

Sets scale operation parameters for

graduation size, creep tracking, zero

tracking, zero range, motion band, digital

filter, overload point, calibration unit,

displayed division size, decimal point

Allows setting and changing of

remote platform parameters

REMOTE

Allows setting and changing of COM

port setups

SETCOM

Sets serial protocol for COM ports

Allows global scale reset to factory

defaults

SETDEF

Resets all scale parameters to factory

defauts

Sets display backlight.

LIGHT

Either always on or conservation mode to

save power for battery operation

TABLE 5-4: Service sub-menu items

5.4.1 CALIB – THREE POINT CALIBRATION

1. Press the soft key CALIB

2. Scale prompts “Select local or remote for calibration”.

3. Press the local or remote soft key or exit soft key to exit calibration

4. The scale prompts “Remove all weights and press enter”.

5. Remove all weight from the scale platform and press the ENT key.

6. The scale prompts “please weight” and briefly displays the zero count value and then

displays “Zero Calibration Done”

7. After computing the zero point the scale prompts “Input test weight in pounds (CAL

Point 1)

8. Enter the test weight value. For CAL point one use a test weight of at least 10% of the

scale capacity and press ENT.

9. The scale prompts “Put test weight on the press enter”

10. Place the test weight on scale and press ENT.

11. The scale briefly displays the count value for CAL point one.

12. Follow the prompts for CAL point 2 and 3. Use approximately 50% of the scale capacity

for CAL 2 and close to 100% of scale capacity for CAL 3.

13. If the calibration was completed successfully, the scale briefly display “Calibration

Passed”

5.4.2 CALIB – SINGLE POINT CALIBRATION

1. Follow steps 1 – 11 in 5.4.1.

2. When the scale prompts “Input test weight in pounds (CAL Point 2) enter 0 the press

ENT

3. The scale prompts “Abort Point 2 CAL?”

4. Press the YES soft key.

5. The scale prompts “Multi point cal aborted” and returns to the service sub-menu.