5.4.3 KcalZR Allows re-entry of the zero counts from last calibration

1. Press the KcalZR soft key.

2. Scale prompts “Select local or remote for calibration”.

3. Press the local or remote soft key or exit soft key to exit calibration

4. Scale displays “Key in Zero Counts XXXXXX “ where XXXXXX is the existing value for

zero counts from the last calibration. The first digit is blinking and will accept new values

entered from the keyboard. Either enter a new value and press ENT or press ENT to

accept the displayed value.

5. Scale prompts “Zero count is:XXXXXX.

6. Press the Yes soft key to accept or No soft key to return to step 4.

5.4.4 KCalSP Allows re-entry of the full scale counts from last calibration

1.

Press the KcalSP soft key

2. Scale prompts “Select local or remote for calibration”.

3. Press the local or remote soft key or exit soft key to exit calibration

4. Scale displays “Key in Full Scale Counts XXXXXX “ where XXXXXX is the existing

value for full scale counts from the last calibration. The first digit is blinking and will

accept new values entered from the keyboard. Either enter a new value and press ENT

or press ENT to accept the displayed value

5. Scale prompts “Full Scale counts is:XXXXXX.

6. Press the Yes soft key to accept or No soft key to return to step 4

5.4.5 LOCAL Sets scale operation parameters for graduation size, creep tracking,

zero tracking, zero range, motion band, digital filter, overload point, calibration unit,

displayed division size, decimal point See Table 5-4.

5.4.6 Remote Sets scale operation parameters for graduation size, creep tracking,

zero tracking, zero range, motion band, digital filter, overload point, calibration unit,

displayed division size, decimal point See Table 5-4

PARAMETER

(Soft Key Text)

DESCRIPTION

AVAILABLE

CHOICES

Graduations

(Grads)

Specifies number of full scale graduations. Value should be

consistent with legal requirements and environmental limits on

the useful system resolution.

1,000 1,500

2,000 2,500

3,000 4,000

5,000 6,000

7,500 8,000

10,000 12,000

15,000 20,000

24,000 30,000

40,000 50,000

100,000

Span

N/A

Creep Tracking

(CrpTrk)

Sets the range within which the scale will attempt to offset the

effects of load cell creep. Use a lower setting for lower capacity

load cells and vice-versa. Selections are in seconds.

Off

1, 3, 5, 10, 15,

20, 30, 50, 60,

120, 180, 240

Zero Track Band

(Ztrack)

Selects the range within which the scale will automatically zero.

Note that the scale must be in standstill to automatically zero.

Selections are in Display Divisions.

0 d, 0.5 d, 1 d

3 d , 5 d