C.1.2 CALIBRATION ERRORS

Message

Explanation

Scale Not Ready

Scale is unstable or there is an internal error in the scale.

ERROR: Value Exceeds Scale

Capacity

Indicates that the test weight value you have chosen

exceeds the scale’s capacity.

ERROR: Value <1% of Scale

Capacity

Indicates that the test weight value you have chosen is

less than 1% of the scale’s capacity.

ERROR: Internal Resolution

Too Low

Occurs when you do not place the test weight on the

scale during calibration. Can also indicate that there is

an internal error in the scale.

Wrong Test Weight

During a multi-point calibration, the entered test weight

value did not match the test weight applied to the scale.

This error can also occur if the following relationship is

not true:
Cal Point 1 < Cal Point 2 < Cal Point 3

Invalid Data

Occurs when you enter a value of zero for the KcalZR

or KcalSP functions. See Chapter 5 for more

information.